Effect of an interactive cardiopulmonary resuscitation assist device with an automated external defibrillator synchronised with a ventilator on the CPR performance of emergency medical service staff: a randomised simulation study

Abstract

Background

The present study evaluates whether the quality of advanced cardiac life support (ALS) is improved with an interactive prototype assist device. This device consists of an automated external defibrillator linked to a ventilator and provides synchronised visual and acoustic instructions for guidance through the ALS algorithm and assistance for face-mask ventilations.

Methods

We compared the cardiopulmonary resuscitation (CPR) quality of emergency medical system (EMS) staff members using the study device or standard equipment in a mannequin simulation study with a prospective, controlled, randomised cross-over study design. Main outcome was the effect of the study device compared to the standard equipment and the effect of the number of prior ALS trainings of the EMS staff on the CPR quality. Data were analysed using analyses of covariance (ANCOVA) and binary logistic regression, accounting for the study design.

Results

In 106 simulations of 56 two-person rescuer teams, the mean hands-off time was 24.5% with study equipment and 23.5% with standard equipment (Difference 1.0% (95% CI: −0.4 to 2.5%); p = 0.156). With both types of equipment, the hands-off time decreased with an increasing cumulative number of previous CPR trainings (p = 0.042). The study equipment reduced the mean time until administration of adrenaline (epinephrine) by 23 s (p = 0.003) and that of amiodarone by 17 s (p = 0.016). It also increased the mean number of changes in the person doing chest compressions (0.6 per simulation; p < 0.001) and decreased the mean number of chest compressions (2.8 per minute; p = 0.022) and the mean number of ventilations (1.8 per minute; p < 0.001). The chance of administering amiodarone at the appropriate time was higher, with an odds ratio of 4.15, with the use of the study equipment CPR.com compared to the standard equipment (p = 0.004). With an increasing number of prior CPR trainings, the time intervals in the ALS algorithm until the defibrillations decreased with standard equipment but increased with the study device.

Conclusions

EMS staff with limited training in CPR profit from guidance through the ALS algorithm by the study device. However, the study device somehow reduced the ALS quality of well-trained rescuers and thus can only be recommended for ALS provider with limited experience.

Background

The quality of out-of-hospital cardiopulmonary resuscitation (CPR) is an important link in the chain of survival and has a direct impact on patient outcomes after out-of-hospital cardiac arrest (OHCA) [1–4]. Nevertheless, both survival to hospital discharge and acceptable neurologic outcome after cardiac arrest remain poor [5, 6]. Although the outstanding importance of the quality of CPR after OHCA has been known for 20 years, the quality of both out-of-hospital and in-hospital resuscitations is often far behind the recommendations of resuscitation guidelines because both emergency medical service (EMS) staff members and in-hospital emergency teams have difficulty with complying with these guidelines [4, 7–12].

According to the 2015 CPR guidelines of the European Resuscitation Council (ERC), rescuers should deliver chest compressions (CC) at a rate of 100–120 compressions per minute, analyse the electrocardiographic (ECG) rhythm as soon as possible, try to change the person doing chest compressions every 2 min, apply ventilations with a compression to ventilation ratio of 30:2 and administer intravenous medication according to cardiac rhythm. Additionally, rescuers should minimise the hands-off time to avoid interrupting CC and should not interrupt CC for more than 10 s. to provide ventilation [3, 13]. Resuscitation guidelines emphasise the importance of limiting the hands-off time to a minimal duration [3, 13].

The hands-off time is due to pausing CC when checking for vital signs, conducting cardiac rhythm analysis, applying bag-mask ventilations, establishing venous access for emergency medication and providing advanced airway management. EMS staff require regular training on the advanced life support (ALS) algorithm, according to resuscitation guidelines [3, 14]. But then, exposure to OHCA is rare, especially in rural areas, and some teams thus have limited experience in ALS [15]. There are various devices commercially available that have been developed to increase the quality of CPR, including metronome, visible and audible feedback systems and active chest compression-decompression assist devices.

The goal of the present study was to evaluate whether the quality of ALS performed by experienced EMS staff could be improved with process assistance from a new interactive prototype device that consists of an automated external defibrillator (AED) linked to a ventilator and provides synchronised visual and acoustic instructions for ALS workflow measures and assistance for face-mask ventilations. We determined the effect of the prototype assist device compared to the standard equipment and the effect of the number of prior ALS trainings of the EMS staff on the CPR quality in a simulation study. The following null hypothesis was tested: the prototype assist device and the number of prior ALS trainings do not affect the CPR performance of the EMS staff in the simulations.

Methods

Study design and participants

We compared the standard CPR equipment (standard equipment) to the interactive device (CPR.com device) in a full-scale mannequin simulation study with a prospective, controlled, randomised cross-over study design. The regional ethics committee of the Medical Council waived the requirements for ethical approval on January 28, 2014.

Data were collected during two standardised simulated cardiac arrest scenarios in three different rescue stations in the Hamburg urban area. Participants were experienced, practicing EMS staff members but had three different levels of vocational education: paramedics, emergency medical technicians (EMTs) and EMTs who were in vocational training to become paramedics. All participants were employees of the G.A.R.D. Ambulance Service in Hamburg, Germany. Recruitment of study participants was on a voluntary basis. There was no selection of volunteering participants. All participants gave written informed consent to participate in the study and evaluate their CPR performance data. All volunteers were regularly trained in ALS for re-certification.

Study protocol

Prior to the study measurements, participants were made familiar with the study protocol, the standard equipment and the CPR.com study device. This education included a short recapitulation of the ALS algorithm of the ERC CPR guidelines, an introduction of the study equipment and instructions on the use of the devices in the study scenario. Participants were instructed to use a bag-valve-mask device or the CPR.com device, to insert a laryngeal tube (LT) in the course of the CPR scenario and to use a mechanical ventilator (standard ventilator or CPR.com device) after securing the airway with a LT.

Each team of two rescuers used the standard equipment in one of the two simulation scenarios and the study equipment in the other. Block randomisation was employed to assign the sequence in which the two treatment scenarios were applied to each rescuer team. A block consisted of four rescuer teams. Immediately prior to the simulation, each team received a sealed envelope containing information on the sequence to be employed.

In all simulation scenarios, participants had to perform ALS on a high-fidelity, full-scale training mannequin (Resusci Anne Simulator, Laerdal Medical AS, Stavanger, Norway) that was placed on the floor, which simulated a patient with OHCA due to persisting ventricular fibrillation (VF).

Each simulation scenario was terminated after the fourth defibrillation shock was administered. The participants had a rest of at least 20 min between the CPR sessions in the study.

Equipment

In all CPR scenarios, the rescuer teams had to use basic equipment, which consisted of tools for establishing peripheral venous access, infusion sets, infusion fluids, syringes, glass ampoules with epinephrine and amiodarone, cannulas, stethoscope, a LT (LTS-D Size 4, VBM Medizintechnik GmbH, Sulz, Germany) for airway management and a portable suction unit. Additionally, participants were randomised to use either the standard or study equipment in each of their CPR scenarios.

Standard equipment

The standard equipment consisted of an AED MEDUCORE Standard (WEINMANN Emergency Medical Technology, Hamburg, Germany) and a ventilator MEDUMAT Standard2 (WEINMANN Emergency Medical Technology) fixed together with a 2-l oxygen bottle on a portable system. The MEDUCORE Standard provided an acoustic metronome of 100/min and announcements prompting the start of CC and rhythm analysis to assist BLS measures.

In the standard equipment CPR session, the MEDUMAT Standard2 was used for ventilation via the inserted LT. At the start of the ventilator in case of a CPR, the participants had to select the type of patient to be ventilated from a menu, which presented the options of an adult patient, an older child or a toddler. After selecting an adult patient, the ventilator applied volume-controlled intermittent positive pressure ventilation (IPPV) with a pre-set frequency of 12/min, a tidal volume (Vt) of 500 ml, a maximum inspiratory pressure (Pmax) of 30 mbar without positive end expiratory pressure (PEEP) and a ratio of inspiration to expiration (I:E) of 1:1.7. Before inserting the LT, the teams provided face-mask ventilations with a bag-valve-mask device (Combibag™, WEINMANN Emergency Medical Technology).

Study equipment (CPR.com)

The CPR.com equipment (WEINMANN Emergency Medical Technology) was a prototype device that was based on the technology of the AED MEDUCORE Standard and the ventilator MEDUMAT Standard2 fixed on a portable system that were linked to and synchronised with each other. CPR.com provided a metronome of 100/min and instructions on BLS and ALS measures, according to the CPR guidelines, both visually on the screen and by voice announcements. These instructions prompted the rescuers to perform ALS measures according to the ERC guidelines and thereby guided BLS, AED use, changeover of the rescuers’ roles and places, airway management, venous access, and the preparation and administration of adrenaline (epinephrine) and amiodarone, if appropriate. After the requested ALS measure was performed, the rescuers confirmed the completion of the ALS measure on the device’s user interface (Fig. 1).

Fig. 1
figure 1

The prototype device CPR.com (WEINMANN Emergency Medical Technology; Hamburg, Germany) with the switch for triggering ventilations via the face mask (MEDUtriggerâ„¢, WEINMANN Emergency Medical Technology) in the foreground and a rescuer communicating with the device

The CPR.com device provided ventilation via a face-mask and a secured airway. Similar to the standard equipment, the rescuers selected the type of patient from a menu. If the participants selected an adult patient, the CPR.com device was automatically adapted to the settings of Vt 500 ml, Pmax 30 mbar, PEEP 0 mbar and I:E 1:1 for face-mask ventilations. These inflations were triggered by a switch at the face-mask (MEDUtrigger), which was connected to the device by a ventilation tube. CPR.com warned the provider in case that the MEDUtrigger had not been used for more than 40 s. After the placement of the LT was confirmed on the device, CPR.com automatically started IPPV with a pre-set frequency of 12/min, a Vt of 500 ml, Pmax 30 mbar, PEEP 0 mbar and I:E of 1:1.7.

The CPR.com device paused ventilations automatically during ECG rhythm analysis and defibrillations.

The rescuers were not familiar with the standard and the study equipment prior to the study.

Data collection

The simulation scenarios were digitally video recorded and analysed for the BLS and ALS measures taken by the participants. The ventilation data were recorded using a differential pressure and flow sensor (Hamilton Medical, Bonaduz, Switzerland) placed between the mannequin’s upper airway and lungs. After completing the two simulations, the study participants were asked to complete a questionnaire on their demographic data, vocational education, CPR experience, prior training in CPR and handling of the CPR.com device.

Outcome measures

The quality of CPR and the degree of CPR guideline compliance of the rescuer teams were measured using prospectively defined primary and secondary outcome variables. The primary endpoint was the hands-off time, which is defined as the percentage of the entire CPR time without CC. Secondary endpoints were the frequency of CC and the total number of CC and ventilations per minute, the number of changes of the person doing CC (rescuer’s change in compressor and ventilator roles) and the proportions of CPRs with adrenaline and amiodarone applied to the correct point in time according to ERC guidelines. Depth and release of CC could not be evaluated, as they are not accurately registered by the software version used in the mannequin when CC and ventilations are applied simultaneously.

Additionally, secondary endpoints were the time intervals until the first, second and third defibrillations, intubation with the LT, implementation of venous access and first administration of adrenaline and amiodarone.

Statistical analysis

The categorical data are presented as counts (percentages) and the continuous data as means with standard deviation (SD), standard error (SE) or 95% confidence interval (CI). We used hierarchical multivariate analyses of covariance (ANCOVA) adapted for a cross-over study design. Binary logistic regression models were used to determine the effect of the study device (CPR.com) on the administration of adrenaline and amiodarone at the correct point in time. All analyses were done using mixed model approaches, reflecting the fact that we have firmly established two-person rescuer teams conducting the same type of investigation two times, each time under a different treatment regime. The rescuer team was therefore considered as random effect. Treatment (equipment), period (to determine carry-over effects) and sequence (i. e. which treatment was administered first) were fixed effects in the models. Cumulative number of prior CPR training sessions attended by either of the team members within the last 5 years were incorporated as covariates in the analyses as well. Additionally, we included the highest vocational education level of the two partners in a rescuer team as a covariate in the model. Sample size calculation was targeted to detect a minimal relevant difference of 5% in the hands-off time, with a power of 80% and a significance level of 5%. The assumed standard deviation of 7% was based on the results of similar studies, and a need for 98 evaluable simulations was found [16, 17]. Based on this number of simulations and accounting for a 12% drop-out rate, the required number of simulations was set at 112.

P-values below 0.05 were considered significant, without correcting for multiple testing. Statistical analysis was performed using the statistical software package R 3.2.3 [18].

Results

One hundred twelve participants were included in the study and formed 56 teams of two ALS providers. According to the crossover design of the study, the teams were randomised to 56 simulation sessions with the study equipment and 56 simulation sessions with standard equipment. Of the 112 consecutive simulations that were initially recorded in the study, a complete data set was analysed from 106 simulations (Fig. 2). The characteristics of the participants in the study are presented in Table 1.

Fig. 2
figure 2

Study flow diagram of allocation to study or standard equipment and data analysis according to CONSORT statement

Table 1 Characteristics of the study participants

In this simulation study, the mean hands-off time as the primary outcome did not differ significantly between the simulations with the study equipment CPR.com and those with the standard equipment (Tables 2 and 3).

Table 2 Quality of CPR achieved with the study device CPR.com and the standard equipment
Table 3 Effects of the use of CPR.com device or standard equipment compared with prior CPR training

The cumulative number of prior CPR trainings attended within the last 5 years by both team members demonstrated a small effect on the hands-off time, i.e., a 0.25% point reduction per prior CPR training of one of the team members (Table 3, Fig. 3). Accordingly, 20 cumulative prior CPR trainings for both team members within the last 5 years (i.e., two trainings per each EMS provider per year) led to a decrease of 5% points in the hands-off time.

Fig. 3
figure 3

Effect plots illustrating the associations between the allocation to the equipment used (CPR.com device vs. standard equipment) and the cumulative number of prior CPR trainings of both team members within the last 5 years and the hands-off time (a), the time until the first defibrillation (b), the time until the second defibrillation (c), the time until the third defibrillation (d), the time interval until the administration of adrenaline (e) and the time interval until the administration of amiodarone (f). Coloured areas represent the estimator’s confidence bands. Although the plots illustrate the observed different effects of the equipment used and the number of prior CPR trainings of the rescuer teams, the interdependency between the equipment and the number of prior CPR trainings only reached statistical significance at the time of the third defibrillation in the ERC algorithm (d) (p = 0.014)

Within the secondary endpoints, the study equipment increased the number of changes of the person doing CC and reduced the time interval until the first administration of adrenaline and amiodarone. However, the use of the study equipment resulted in a reduced number of CC and ventilations per minute compared to the standard equipment (Tables 2 and 3).

Furthermore, the CPR.com equipment affected the time that the participants needed to pass through the ERC algorithm until the third defibrillation, but this effect was subject to a significant interaction with the cumulative number of prior CPR trainings (p = 0.014, Fig. 3). Figure 3 illustrates the association among the equipment used, the number of prior CPR trainings of the rescuer teams, the hands-off time and the time intervals that the participants needed to pass through the ERC algorithm.

In the simulation sessions with the standard equipment, the time until the first, second and third defibrillations and that until the administration of adrenaline and amiodarone decreased with the number of prior CPR trainings of the participants. In contrast, in the simulation sessions with the study device, the time until the defibrillations increased with the number of prior CPR trainings of the participants. Those teams with only two cumulative CPR trainings in the last five years needed shorter times until defibrillation with the study device than did the teams using the standard equipment (Fig. 3). However, those teams with 18 cumulative CPR trainings in the last five years needed longer times until the defibrillations with the study device compared with the standard equipment (Fig. 3). The analyses showed that teams were faster in the ERC algorithm until the first and third defibrillations with the study device compared to the standard equipment unless they had attended more than approximately 10 cumulative ALS trainings within the last five years per two-person team, which corresponds to one training per participant per year.

Logistic regression revealed that the chance of administering amiodarone at the appropriate time according to the ERC guidelines was approximately four times higher with the use of the study equipment CPR.com compared to the standard equipment (p = 0.004). The chance of administering adrenaline at the appropriate time showed an advantage for the CPR.com device as well, but the results did not differ significantly between the study device and the standard equipment (p = 0.386) (Fig. 4).

Fig. 4
figure 4

Forest plot of the administration of adrenaline and amiodarone at the correct time according to ERC guidelines as determined with binary logistic regression analyses accounting for cross-over design

There was no significant effect of the level of vocational education of the participants on the quality of CPR.

A comprehensive dataset with the data analysed in the study is available from the corresponding author on request.

Discussion

In this simulation study, an interactive prototype guiding defibrillation and ventilation in ALS was used safely by experienced EMS staff after a short introduction to the device.

The participants performed high-quality CPR with both the study and standard equipment compared to other study results. The hands-off times of 24.5 and 23.5%, respectively, were within the same range of the hands-off times reported in other studies (between 16 and 30%) [5, 10, 19–21]. In this study, the new prototype device was not superior to the standard equipment with respect to the hands-off time, whereas the study device slightly decreased the number of CC and ventilations per minute. However, although the differences between the study and standard equipment showed statistically significant effects, these effects are supposed to be of little clinical relevance to the patient’s outcome. The cause for failing to prove superiority might be the lack of familiarity with the equipment and the complexity of the scenario. A previous study by Rittenberger and colleagues showed that EMS focus not enough on BLS in a complex resuscitation scenario [22]. It remains unclear, whether the observed significant differences gain clinical importance in routine use. Some previous studies have found a positive association between the chest compression fraction (CCF) and survival, whereas others found diverging or inconsistent results and concluded that CCF alone is not a predictor of survival of OHCA [1, 21, 23–25]. In the present study, the hands-off time decreased with the number of prior CPR trainings of the participants but did not change with the use of the study device. There is evidence from clinical studies that the greatest chance of survival after OHCA is with CC rates between 100 and 120/min, which the participants achieved with both types of equipment [2, 26].

The increased number of changes of the person doing chest compressions of 0.6 per CPR session might have contributed to longer hands-off times. However, the ERC guidelines recommend changing the person doing chest compressions every 2 min to maintain high-quality CPR, and therefore, the higher number of rescuer’s changeovers might be of clinical relevance [3, 13].

The interaction with the device, which requested performing actions and confirming their completion, may have required time and intention that the EMS staff had available for CC while using the standard equipment. It is unknown whether the results would have been the same if the EMS staff were as familiar with the study device as they were with the standard equipment, which most of them use in daily practice.

In regard to the secondary endpoints, we found a significant interaction between the effect of the CPR.com device and the number of prior CPR trainings in the two-rescuer teams. As shown in Fig. 3, there was an association between the times that the participants needed to work through the ALS algorithm of the ERC, the allocation to the use of the CPR.com device or standard equipment and the cumulative number of prior CPR trainings of both team members within the last 5 years. Participants with fewer prior CPR trainings were faster with the study device than with standard equipment, whereas well-trained participants were possibly slowed down by the study device compared with the standard equipment. These well-trained rescuers needed more time to work through the algorithm with the study device than with the standard equipment. We refer these findings to the fact that by using the study equipment ALS measures were performed only after the CPR.com device prompted them and that the participants had to confirm the completion of an ALS measure on the device’s user interface. Thereby the need to interact with the study device extended the time that the rescuers needed to work through the ALS algorithm. On the other hand, an interactive device guiding the rescuer team through the ALS algorithm is of major help for rescuers who have had little regular training [17].

The time until the administration of adrenaline was shorter with the CPR.com device than with standard equipment, but it was not proven that adrenaline was administered at the correct time more frequently according to the ERC algorithm. This effect can be explained by the study device prompting the participants to prepare adrenaline for iv-injection in advance of the correct time for adrenaline injection. In some cases, this instruction led the participants to administer adrenaline before the CPR.com device indicated the correct time for the adrenaline injection. A better-timed adrenaline injection might be achieved if the device is used regularly. The present study device was designed to guide the rescuer through the ALS algorithm of current guidelines.

However, our results support the recommendations of the 2015 ERC guidelines on the use of prompt devices, which is to use CPR prompt devices only as part of a broader system of CPR quality improvement initiatives rather than as an isolated intervention [3, 13].

The present study has some limitations. First, the CPR quality was evaluated on a mannequin and not in real-life routine patient care with prolonged resuscitation efforts. As in every simulation study, the standardised study setting tries to control for confounders and thereby increases the reliability and validity of the results. On the other hand, the standardised study setting limits transferability of the results into clinical practice. Second, the depth and release of CC and the ventilation’s tidal volumes were not analysed in this simulation study, because according to the manufacturer of the mannequin these data were not validly measured by the software version used in the study when performing simultaneous CC and ventilations as recommended during ALS. A further limitation of the study is the fact that ventilator settings in both, study and standard equipment, were set to standard ventilator settings of critically ill patients. Although the pre-set tidal volume of 500 ml corresponds well to resuscitation guidelines, the ventilation frequency of 12/min used in the study differs from ERC resuscitation guidelines, which recommend 10 breaths/min [13]. Nevertheless, these settings can easily be adjusted in the study device as well as in the standard equipment to better comply with resuscitation guidelines.

Additionally, although the EMS staff members included in this study were professional rescuers working in an urban area with a high frequency of patient care with CPR, they had a mean of only 4 years of professional experience and thereby the participants might be considered inexperienced. Besides, some participants could only report a limited number of CPR trainings within the last 5 years with a mean of 4 years of professional experience in EMS. However, the CPR performance by EMS staff with comparable experience has been evaluated successfully in a similar simulation study and is supposed to be the group profiting the most from such an interactive device [27].

Conclusion

The study device CPR.com improved the ALS quality of EMS staff with a limited number of prior ALS trainings, but it somehow reduced the ALS quality of providers who were well trained in ALS. Given the limitations of a simulation study, the interactive study device can only be recommended for less trained rescuers and not for rescuers who are training in and performing CPR on a regular basis. EMS crew members may potentially benefit from this assist device, if they have limited experience in CPR, for example during their professional training, or if they have a low frequency of CPR performance, for example in rural areas.

Abbreviations

AED:

Automated external defibrillator

ALS:

Advanced life support

ANCOVA:

Analyses of covariance

CC:

Chest compressions

CI:

Confidence interval

CPR:

Cardiopulmonary resuscitation

ECG:

Electrocardiography

EMS:

Emergency medical system

EMT:

Emergency medical technician

ERC:

European Resuscitation Council

IPPV:

Intermittent positive pressure ventilation

LT:

Laryngeal tube

OHCA:

Out-of-hospital cardiac arrest

PEEP:

Positive end expiratory pressure

Pmax:

Maximum inspiratory pressure

SD:

Standard deviation

SE:

Standard error

VF:

Ventricular fibrillation

Vt:

Tidal volume

References

  1. Christenson J, Andrusiek D, Everson-Stewart S, Kudenchuk P, Hostler D, Powell J, et al. Chest compression fraction determines survival in patients with out-of-hospital ventricular fibrillation. Circulation. 2009;120:1241–7.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Idris AH, Guffey D, Aufderheide TP, Brown S, Morrison LJ, Nichols P, et al. Relationship between chest compression rates and outcomes from cardiac arrest. Circulation. 2012;125:3004–12.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Perkins GD, Handley AJ, Koster RW, Castren M, Smyth MA, Olasveengen T, et al. European resuscitation council guidelines for resuscitation 2015: Section 2. Adult basic life support and automated external defibrillation. Resuscitation. 2015;95:81–99.

    Article  PubMed  Google Scholar 

  4. Ko PC, Chen WJ, Lin CH, Ma MH, Lin FY. Evaluating the quality of prehospital cardiopulmonary resuscitation by reviewing automated external defibrillator records and survival for out-of-hospital witnessed arrests. Resuscitation. 2005;64:163–9.

    Article  PubMed  Google Scholar 

  5. Olasveengen TM, Wik L, Steen PA. Quality of cardiopulmonary resuscitation before and during transport in out-of-hospital cardiac arrest. Resuscitation. 2008;76:185–90.

    Article  PubMed  Google Scholar 

  6. Berdowski J, Berg RA, Tijssen JG, Koster RW. Global incidences of out-of-hospital cardiac arrest and survival rates: Systematic review of 67 prospective studies. Resuscitation. 2010;81:1479–87.

    Article  PubMed  Google Scholar 

  7. Gallagher EJ, Lombardi G, Gennis P. Effectiveness of bystander cardiopulmonary resuscitation and survival following out-of-hospital cardiac arrest. JAMA. 1995;274:1922–5.

    Article  CAS  PubMed  Google Scholar 

  8. Valenzuela TD, Kern KB, Clark LL, Berg RA, Berg MD, Berg DD, et al. Interruptions of chest compressions during emergency medical systems resuscitation. Circulation. 2005;112:1259–65.

    Article  PubMed  Google Scholar 

  9. Abella BS, Sandbo N, Vassilatos P, Alvarado JP, O’Hearn N, Wigder HN, et al. Chest compression rates during cardiopulmonary resuscitation are suboptimal: a prospective study during in-hospital cardiac arrest. Circulation. 2005;111:428–34.

    Article  PubMed  Google Scholar 

  10. Abella BS, Alvarado JP, Myklebust H, Edelson DP, Barry A, O’Hearn N, et al. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. JAMA. 2005;293:305–10.

    Article  CAS  PubMed  Google Scholar 

  11. Wik L, Kramer-Johansen J, Myklebust H, Sorebo H, Svensson L, Fellows B, et al. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. JAMA. 2005;293:299–304.

    Article  CAS  PubMed  Google Scholar 

  12. Kramer-Johansen J, Myklebust H, Wik L, Fellows B, Svensson L, Sorebo H, et al. Quality of out-of-hospital cardiopulmonary resuscitation with real time automated feedback: a prospective interventional study. Resuscitation. 2006;71:283–92.

    Article  PubMed  Google Scholar 

  13. Soar J, Nolan JP, Bottiger BW, Perkins GD, Lott C, Carli P, et al. European resuscitation council guidelines for resuscitation 2015: Section 3. Adult advanced life support. Resuscitation. 2015;95:100–47.

    Article  PubMed  Google Scholar 

  14. Greif R, Lockey AS, Conaghan P, Lippert A, De Vries W, Monsieurs KG, et al. European resuscitation council guidelines for resuscitation 2015: Section 10. Education and implementation of resuscitation. Resuscitation. 2015;95:288–301.

    Article  PubMed  Google Scholar 

  15. Dyson K, Bray J, Smith K, Bernard S, Straney L, Finn J. Paramedic exposure to out-of-hospital cardiac arrest is rare and declining in Victoria, Australia. Resuscitation. 2015;89:93–8.

    Article  PubMed  Google Scholar 

  16. Lukas RP, Engel P, Wecker S, Thies S, Friederichs H, Gerss J, et al. Cardiopulmonary resuscitation guidance improves medical students’ adherence to guidelines in simulated cardiac arrest: a randomised cross-over study. Eur J Anaesthesiol. 2013;30:752–7.

    Article  PubMed  Google Scholar 

  17. Busch CW, Qalanawi M, Kersten JF, Kalwa TJ, Scotti NA, Reip W, et al. Providers with limited experience perform better in advanced life support with assistance using an interactive device with an automated external defibrillator linked to a ventilator. J Emerg Med. 2015;49:455–63.

    Article  PubMed  Google Scholar 

  18. R Core Team. R: A language and environment for statistical computing. Vienna, Austria: Foundation for Statistical Computing; 2015. URL https://www.R-project.org/.

    Google Scholar 

  19. Bohn A, Weber TP, Wecker S, Harding U, Osada N, Van Aken H, et al. The addition of voice prompts to audiovisual feedback and debriefing does not modify CPR quality or outcomes in out of hospital cardiac arrest--a prospective, randomized trial. Resuscitation. 2011;82:257–62.

    Article  PubMed  Google Scholar 

  20. Muller MP, Richter T, Papkalla N, Poenicke C, Herkner C, Osmers A, et al. Effects of a mandatory basic life support training programme on the no-flow fraction during in-hospital cardiac resuscitation: an observational study. Resuscitation. 2014;85:874–8.

    Article  PubMed  Google Scholar 

  21. Nichol G, Leroux B, Wang H, Callaway CW, Sopko G, Weisfeldt M, et al. Trial of continuous or interrupted chest compressions during CPR. N Engl J Med. 2015;373:2203–14.

    Article  CAS  PubMed  Google Scholar 

  22. Rittenberger JC, Guimond G, Platt TE, Hostler D. Quality of BLS decreases with increasing resuscitation complexity. Resuscitation. 2006;68:365–9.

    Article  PubMed  Google Scholar 

  23. Vaillancourt C, Everson-Stewart S, Christenson J, Andrusiek D, Powell J, Nichol G, et al. The impact of increased chest compression fraction on return of spontaneous circulation for out-of-hospital cardiac arrest patients not in ventricular fibrillation. Resuscitation. 2011;82:1501–7.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Rea T, Olsufka M, Yin L, Maynard C, Cobb L. The relationship between chest compression fraction and outcome from ventricular fibrillation arrests in prolonged resuscitations. Resuscitation. 2014;85:879–84.

    Article  PubMed  Google Scholar 

  25. Cheskes S, Schmicker RH, Rea T, Powell J, Drennan IR, Kudenchuk P, et al. Chest compression fraction: a time dependent variable of survival in shockable out-of-hospital cardiac arrest. Resuscitation. 2015;97:129–35.

    Article  PubMed  Google Scholar 

  26. Idris AH, Guffey D, Pepe PE, Brown SP, Brooks SC, Callawayet CW, et al. Chest compression rates and survival following out-of-hospital cardiac arrest. Crit Care Med. 2015;43:840–8.

    Article  PubMed  Google Scholar 

  27. Aufderheide TP, Pirrallo RG, Yannopoulos D, Klein JP, von Briesen C, Sparks CW, et al. Incomplete chest wall decompression: a clinical evaluation of CPR performance by EMS personnel and assessment of alternative manual chest compression–decompression techniques. Resuscitation. 2005;64:353–62.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank all EMS staff members of the G.A.R.D. Ambulance Service in Hamburg, Germany for their enthusiastic support and participation in the study.

Funding

The conduct of the study was financially supported by a grant from the Hamburg Innovation Foundation (grant number: IT-721.230-002 (466)). The Foundation had no influence on the study design; on the collection, analysis or interpretation of the data; or on the writing of the manuscript.

Availability of data and materials

The complete dataset supporting the conclusions of this article is available from the corresponding author on request.

Authors’ contributions

All authors together designed the trial. The authors RN, CD, JB, NAS and TJK developed the CPR assist device and conceived the study. CD, JB and NAS recruited the study participants and managed the data. CD, JB and NAS collected the data, and RN and JCK supervised the conduct of the trial and data collection including internal validation. RN and JFK provided statistical advice on the study design and analysed the data. RN and CD drafted the manuscript, and all authors contributed substantially to its revision. All authors approved the final version to be submitted. RN takes responsibility for the paper as a whole.

Competing interests

The authors RN, CD, JFK and JCK declare no competing interests or personal relationships with people or organisations that could influence their work. The authors TJK and NAS are full-time employees on a fixed salary at WEINMANN Emergency Medical Technology, Hamburg, Germany. The prototype and all patent rights are owned by WEINMANN Emergency. None of the authors hold any patent rights on the prototype device, and none of the authors are involved in sales profits.

Consent for publication

Written informed consent for the publication from the participants enrolled in this study may be provided on request.

Ethics approval and consent to participate

The study protocol was submitted to the regional ethics committee of the Medical Council, which waived the need for ethical approval on January 28, 2014. All participants were informed about the purpose of the study and gave written informed consent.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Rainer Nitzschke.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nitzschke, R., Doehn, C., Kersten, J.F. et al. Effect of an interactive cardiopulmonary resuscitation assist device with an automated external defibrillator synchronised with a ventilator on the CPR performance of emergency medical service staff: a randomised simulation study. Scand J Trauma Resusc Emerg Med 25, 36 (2017). https://doi.org/10.1186/s13049-017-0379-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13049-017-0379-8

Keywords