Skip to main content

Table 4 BLS with the use of an AED in scenarios with persistent asystole: optimized sequence of actions

From: Quality of basic life support when using different commercially available public access defibrillators

 

Time until 1st chest compression [seconds]

Time until completion of 1st rhythm analysis [seconds]

Duration of rhythm analysis [seconds]

Interval between completion of 1st rhythm analysis and start of 2nd rhythm analysis [seconds]

No-flow fraction

(A) Heartsave PAD

40 ± 2C,E,+

39 ± 3

19 ± 2

140 ± 1B

0.41 ± 0.01

(B) Philips HS1

26 ± 1D,G,H,K,+

24 ± 1D,G,H,+

10 ± 1C,D,G,H

142 ± 1A

0.33 ± 0.02C,D,E,F,G,H,+

(C) Lifeline VIEW

33 ± 2A,D,E,G,H,+

32 ± 2D,E,+

12 ± 1B,D,E,F

153 ± 0

0.36 ± 0.02B,D,E,F,G,H,+

(D) Lifepak CR Plus

28 ± 2B,C,E,G,H,K,+

26 ± 1B,C,E,G,H,+

10 ± 1B,C,G,H

125 ± 0G

0.33 ± 0.01B,C,E,F,G,H

(E) PowerHeart G5

34 ± 0A,C,D,G,+

32 ± 1C,D,+

14 ± 1C,F

184 ± 1

0.34 ± 0.01B,C,D,F,G,H,+

(F) Fred Easy Life

16 ± 2+

48 ± 1+

15 ± 1C,E

122 ± 1G

0.33 ± 0.01B,C,D,E,G,H,+

(G) AED Plus

27 ± 2B,C,D,E,H,K,+

26 ± 1B,D,H,+

9 ± 0B,D,H

125 ± 1D,F

0.35 ± 0B,C,D,E,F,H,+

(H) Cardiolife AED 2100

27 ± 3B,C,D,G,K,+

25 ± 3B,D,G,+

9 ± 0B,D,G

128 ± 1

0.35 ± 0B,C,D,E,F,G

(K) BLS without AED

25 ± 2B,D,G,H

   

0.26 ± 0.01

  1. The duration of the scenario was 300 seconds. Values are the mean ± standard deviation. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with a standard sequence of actions.