Skip to main content

Table 3 BLS with the use of an AED in scenarios of persistent ventricular fibrillation: optimized sequence of actionsThe duration of the scenario was 300 seconds

From: Quality of basic life support when using different commercially available public access defibrillators

 

Time until 1st chest compression [seconds]

Time until completion of 1st rhythm analysis [seconds]

Duration of rhythm analysis [seconds]

Time until 1st shock [seconds]

Interval between 1st shock and start of 2nd rhythm analysis [seconds]

Perishock pause [seconds]

No-flow fraction

(A) Heartsave PAD

35 ± 2B,C,E,G

34 ± 2 B,C,E

14 ± 1 B,C,E,F

47 ± 2 F

137 ± 1 B

3 ± 1 D,+

0.36 ± 0.01B,C,D,E,F,G,H,+

(B) Philips HS1

31 ± 2A,D,E,G,H,K,+

28 ± 2 A,D,E,G,H,+

14 ± 2 A,C,E,F,G

30 ± 2 E,G,H,+

140 ± 0 A

10 ± 1 E,H,+

0.34 ± 0.01A,C,D,E,F,G,H,+

(C) Lifeline VIEW

41 ± 1 A,+

35 ± 1 A,E,+

11 ± 2 A,B,D,E,G,H

39 ± 1 D,+

154 ± 1

18 ± 1 F,G,+

0.36 ± 0.01A,B,D,E,F,G,H,+

(D) Lifepak CR Plus

28 ± 2B,E,G,H,K,+

27 ± 2 B,E,G,H,+

10 ± 1 B,C,E,G,H

38 ± 3 C,E,+

125 ± 1 F,G,H

1 ± 1 A,+

0.32 ± 0.01A,B,C,E,F,H,+

(E) PowerHeart G5

34 ± 1A,B,D,G,H,+

30 ± 2 A,B,C,D,G,+

12 ± 1 A,B,C,D,F,G,H

32 ± 1 B,D,G,H,+

185 ± 2

13 ± 2 B,G,H,+

0.33 ± 0.01A,B,C,D,F,G,H,+

(F) Fred Easy Life

17 ± 2+

49 ± 1+

15 ± 1 A,B,E

51 ± 2 A,+

124 ± 1 D,G,H

19 ± 1 C,G,+

0.35 ± 0.01A,B,C,D,E,G,H,+

(G) AED Plus

33 ± 2A,B,D,E,H,+

27 ± 2 B,D,E,H,+

10 ± 1 B,C,D,E,H

31 ± 2 B,E,H,+

125 ± 1 D,F,H

18 ± 1 C,E,F,+

0.37 ± 0.02A,B,C,E,F,H

(H) Cardiolife AED 2100

27 ± 1B,D,E,G,K,+

23 ± 0 B,D,G,+

8 ± 1 C,D,E,G

26 ± 1 B,E,G,+

127 ± 2 D,F,G

11 ± 1 B,E

0.35 ± 0.02A,B,C,D,E,F,G

(K) BLS without AED

25 ± 2B,D,H

     

0.26 ± 0.01

  1. Values are the mean ± standard deviation. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with a standard sequence of actions