Table 2 BLS with the use of an AED in scenarios of persistent asystole

From: Quality of basic life support when using different commercially available public access defibrillators

 

Time until 1st chest compression [seconds]

Time until completion of 1st rhythm analysis [seconds]

Duration of rhythm analysis [seconds]

Interval between completion of 1st rhythm analysis and start of 2nd rhythm analysis [seconds]

No-flow fraction

(A) Heartsave PAD

58+ ± 2 D,E,H,I,+

50 ± 3D,H,I

20 ± 1

140 ± 1C

0.52 ± 0.03C

(B) Philips HS1§

112 ± 2G

77 ± 3C,G

11 ± 1C,D,E,I

200 ± 1

0.71 ± 0F

(C) Philips HS1*

92 ± 2G,+

78 ± 2B,G,+

12 ± 0B,D,E,I

140 ± 0A

0.50 ± 0A,+

(D) Lifeline VIEW

64 ± 1A,E,H,I,+

53 ± 2A,H

12 ± 1B,C,F,I

154 ± 1

0.58 ± 0.01+

(E) Lifepak CR Plus

67 ± 1 A,D,H,I,+

61 ± 1+

10 ± 1B,C,H,I

125 ± 0G,H,I

0.42 ± 0.01G,H,I

(F) PowerHeart G5

148 ± 13+

96 ± 2+

15 ± 1D,G

183 ± 1

0.72 ± 0B,+

(G) Fred Easy Life

97 ± 1 B,C,+

74 ± 1B,C,+

15 ± 0F

123 ± 1E

0.45 ± 0E,H,+

(H) AED Plus

57 ± 1A,D,E,I,+

53 ± 2A,D,+

8 ± 0E,I

126 ± 1E,I

0.41 ± 0.01E,G,I,+

(I) Cardiolife AED 2100

52 ± 1A,D,E,H,+

45 ± 1A,+

10 ± 0B,C,D,E,H

127 ± 0E,H

0.40 ± 0.01E,H

(K) BLS without AED

25 ± 2

   

0.26 ± 0.01

  1. The duration of the scenario was 300 seconds. Values are the mean ± standard deviation. The Philips HS1 asks the rescuer to press an “info button” for detailed information about chest compressions and ventilations after each rhythm analysis. The scenario was run 3 times when pressing the info button§ and 3 times without pressing the button*. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with an optimized sequence of actions.