Skip to main content

Table 2 Overview of the arterial pressure variables measured by the oscillometric method and vascular unloading technique and the differences between the two methods

From: Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department

  Mean value OSCI (± SD) Mean value VUT (± SD) Mean difference (± SD; LOA)
SAP 130 (± 24) 125 (± 27) -5 (± 22; -47 to + 37)
DAP 74 (± 17) 71 (± 15) -2 (± 15; -32 to + 27)
MAP 97 (± 20) 92 (± 19) -6 (± 16; -37 to + 26)
  1. All arterial pressure values are given in mmHg. OSCI, oscillometric arterial pressure measurement; SD, standard deviation; SAP, systolic arterial pressure; DAP, diastolic arterial pressure; MAP, mean arterial pressure; VUT, vascular unloading technique; LOA, limits of agreement.