Table 1 Accuracy of syringe infusion pump devices in delivering a prespecified infusion flow rate set at 10 mL/h depending on altitude using weight measurement method [Method A])*

From: Accuracy of low-weight versus standard syringe infusion pump devices depending on altitude

  Syringe infusion pump device
DPS Orchestra® Agilia Injectomat® Micrel Micropump™ 1 Micrel Micropump™ 2 Micrel Micropump™ 3
Altitude 300 m
 Median 9.78 9.78 9.78 9.72 9.70
  (IQR) (9.57–9.88) (9.62–9.84) (9.62–9.88) (9.43–9.79) (9.43–9.79)
 Min–Max 9.13–9.97 9.02–10.04 8.57–10.04 8.48–9.96 8.47–10.10
Altitude 1700 m
 Median 9.66 9.76 9.69 9.66 9.80
  (IQR) (9.54–9.75) (9.62–9.85) (9.45–9.81) (9.36–9.80) (9.62–9.90)
 Min–Max 8.64–9.92 9.38–10.01 8.17–9.94 9.20–9.92 7.49–10.33
  1. IQR Inter-Quartile Range. Median infusion flow rates (p = 0.83) and interquartile ranges (p = 0.27) did not differ across syringe infusion pump devices. Median infusion flow rates (p = 0.15) and interquartile ranges (p = 0.71) did not differ depending on altitude. Median infusion flow rates (p for interaction =0.32) and interquartile ranges (p for interaction =0.66) for syringe infusion pumps did not vary significantly with altitude