Peripheral measurements of venous oxygen saturation and lactate as a less invasive alternative for hemodynamic monitoring

Abstract

Background

Peripheral measurement of venous oxygen saturation and lactate is a less invasive alternative to monitor tissue oxygenation as compared to measurements from a central venous catheter. However, there is a lack of evidence to support the use of peripheral measurements. In this study, we investigated the agreement between central and peripheral venous oxygen saturation and lactate.

Methods

We conducted a prospective observational study including 115 patients who underwent elective cardiac surgery between April and May 2015 at Rigshospitalet, Copenhagen, Denmark. Measurements were obtained simultaneously at induction of anaesthesia, upon arrival in the ICU and 3–4 h postoperatively. Bias and trending ability was identified using Bland-Altman analysis and a four-quadrant plot.

Results

Bias was 13.37% for venous oxygen saturation preoperatively (95% CI: 11.52–15.22, LoA: ±19.10, PE: 22.08%), 11.29% at arrival to the ICU (95% CI: 8.81–13.77, LoA: ±25.10, PE: 32.39%) and 16.49% at 3–4 h postoperatively (95% CI: 14.16–18.82, LoA: ±21.20, PE: 26.82%). A four-quadrant plot demonstrated an 89% concordance. Central and peripheral lactate showed a bias of 0.14 mmol/L preoperatively (95% CI: 0.11–0.17, LoA: ±0.30, PE: 32.08%), 0.16 mmol/L at arrival to ICU (95% CI: 0.09–0.23, LoA: ±0.70, PE: 38.88%) and 0.23 mmol/L at 3–4 h postoperatively (95% CI: 0.11–0.35, LoA: ±0.50, 25.18%).

Discussion

Measurements of peripheral oxygen saturation and lactate may be valuable in an emergency setting, avoiding unnecessary and time consuming application of a CVC.

Conclusion

We found a high bias but an acceptable trending ability between central and peripheral venous oxygenation. Central and peripheral lactate had excellent agreement. Further studies are necessary to validate the use of peripheral venous samples to identify patients at risk of impaired tissue oxygenation.

Background

Adequate tissue oxygenation is necessary to maintain organ function. Venous oxygen saturation reflects tissue oxygenation necessary to maintain normal organ function and is dependent on cardiac output (CO), oxygen concentration of arterial blood (SaO2), oxygen carrying capacity (Hb) and metabolic demands of the body (VO2) according to Fick’s equation [1].

Low venous oxygen saturation reflects impaired tissue oxygenation and is routinely used as a predictive tool and a therapeutic target to improve the care of critically ill patients [2]. Previous studies have associated impaired venous oxygen saturation with increased mortality and morbidity in both surgical patients [3,4,5,6,7,8,9] and patients admitted in the Intensive Care Unit (ICU) [10,11,12,13,14,15,16,17,18,19,20].

Venous oxygen saturation is most commonly measured as central (ScvO2) or mixed (SmvO2) venous oxygen saturation obtained from a central venous catheter (CVC) or a pulmonary artery catheter, respectively [14]. Measurement of venous oxygen saturation (SpvO2) from a peripheral venous cannula is a less invasive alternative to identify patients at risk and help manage patients without a central or a pulmonary artery catheter. However, there is a lack of evidence to support the use of SpvO2. A recent study found a moderate agreement between ScvO2 and SpvO2 measured in patients admitted to the Emergency Department and the ICU. Furthermore, they found central and peripheral lactate levels to have excellent agreement [21]. Only one study has investigated the role of SpvO2 as a substitute for ScvO2, which has not been assessed in a non-emergent clinical settings before.

We conducted a prospective observational study to compare measurements of venous oxygen saturation and lactate in blood samples obtained from a central and a peripheral venous cannula in patients admitted for elective cardiac surgery. In this study, we hypothesize that peripheral venous sampling can estimate measurements of central venous oxygen saturation and lactate, as a less invasive alternative to measure patient hemodynamic.

Methods

Study design

The study was conducted as a prospective single-centre observational study. Patients who underwent elective cardiac surgery, from April 29, 2015 to May 16, 2015 at Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark, were included. Patients with a central and a peripheral venous cannula were eligible for the study. Exclusion criteria were age below 18 years and acute surgical patients.

The collection and processing of data were approved by the Danish Data and Protection Agency (j. no: 2012-58-0004). The regional ethics commission of the Copenhagen Region waived the need for informed consent as the study was appreciated as a quality-improvement study.

Patient management

Surgery was performed through a median sternotomy with patients on cardiopulmonary bypass (CPB). Anaesthesia was induced with propofol 1–2 mg/kg i.v. and fentanyl 5 μg/kg i.v., and maintained with sevoflurane 0.5–3% and remifentanil infusion 0.3–.06 mcg/kg/hour. All patients were intubated and mechanically ventilated with 8 ml/kg, frequency 10–12/min, FiO2 0.6, PEEP 5–10 cmH2O, pCO2 4.5–6.5 kPa. Perioperative treatment of transfusions, fluid or drug infusion was given according to the National Guidelines and routine management at the department.

Study protocol

The CVC was placed after anaesthesia induction. Routinely a second peripheral cannula was inserted simultaneously with placement of the CVC. The peripheral intravenous site was determined by the treating physician with no input from the study investigators. Venous blood samples were obtained simultaneously from the central and peripheral venous cannula, in the latter after application of a tourniquet and with a maximum of 5 min between the two samples. The peripheral venous cannula site and tourniquet application time was registered. Any fluid or drug infusion was discontinued during blood sampling. Approximately 0.2 mL of blood was collected for analysis of gases after an adequate waste draw of 5–10 mL. Blood samples were obtained at pre-specified time-points: as soon as the CVC was inserted, upon arrival in the ICU and 3–4 h postoperatively. All blood samples were analysed on an ABL 800 FLEX radiometer. Results from the central venous blood gases were available to the treating physician.

Statistics

A Bland-Altman plot was performed to describe the agreement between the two methods at all three time points. The bias and 95% limits of agreement (LoA) with confidence interval (CI) and the percentage error (PE) calculated according to Bland-Altman for all measurements [22]. The PE was calculated as 1.96 times SD of the bias divided by the mean value of ScvO2 and SpvO2. The results were categorized according to previous studies defining the agreement between two measurements using Bland-Altman plots [23, 24].

In a sub analysis the preoperative data were divided in to two subgroups (SpvO2>/< 93%) to analyse the arterialisation of SpvO2 observed in some patients. A four-quadrant plot was constructed to assess the trending ability.

Statistical analyses were performed using SPSS version 22.

Results

A total of 115 patients were included in the study. The majority of patients underwent coronary artery bypass grafting (50%), valve replacement (26%) and combined coronary artery bypass grafting and valve replacement (11%). The remaining patients underwent other cardiac surgical procedures including myectomy and repair of aortic aneurism (13%). Among the peripheral blood samples, regardless of the time point, 9% were obtained with the use of a tourniquet whereof 4% had a tourniquet time above 5 s. The majority of peripheral samples were obtained from a dorsal vein on the hand (72%) or from a cubital vein (28%). A total of 316 paired measurements of central and venous oxygen saturation and lactate were obtained from three different sampling points (preoperative, upon arrival to the ICU and 3–4 h postoperative).

Venous oxygen saturation

The mean value of ScvO2 and SpvO2 was 80% (±7) and 93 (±8) preoperatively, 72% (±7) and 83% (±13) on arrival to the ICU and 71% (±7.5) and 87% (±10) at 3–4 h postoperatively, respectively. Venous oxygen saturation was on average higher when obtained from a dorsal hand vein compared to cubital samples (Table 1).

Table 1 Venous oxygen saturation at different sampling locations

The Bland-Altman plots for venous oxygen saturation are presented in Fig. 1ac. Bias was 13.37% in the preoperative measurements (95% CI: 11.52–15.22, LoA: ±19.10), 11.29% at arrival to the ICU (95% CI: 8.81–13.77, LoA: ±25.10) and 16.49% at 3–4 h postoperatively (95% CI: 14.16–18.82, LoA: ±21.20). Preoperatively at baseline, 50% of the blood samples had a SpvO2 > 93%, indicating arterialisation of blood. For this reason, a sub analysis with the Bland-Altman plot was performed in two separate groups: (SpvO2 >/< 93%). Bias was 17.20% (95% CI: 15.80–18.59, LoA: ±11.90) and 4.94% (95% CI: 1.09–8.79, LoA: ±21.60) in the high and the low preoperative group, respectively. The percentage error was 22.08% preoperatively, 32.39% at arrival to the ICU group and 26.83% at 3–4 h postoperatively. The linear regression line was R2 =0.12.

Fig. 1
figure 1

a Peripheral and central venous oxygen saturation paired values, differences against means preoperatively. The Bland-Altman plot presents the differences against the means of peripheral and central venous oxygen saturation measurements obtained preoperatively. A total of 114 paired measurements were obtained. Bias was 13.37 (95% CI: 11.52–15.22, LoA: 19.10). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 22.08%. Abbreviations: SpvO2: peripheral venous oxygen saturation. ScvO2: central venous oxygen saturation. b Peripheral and central venous oxygen saturation paired values, differences against means at arrival to the ICU. The Bland-Altman plot presents the differences against the means of peripheral and central venous oxygen saturation measurements obtained at arrival to the ICU following surgery. A total of 110 paired measurements were obtained. Bias was 11.29 (95% CI: 8.81–13.77, LoA: 25.10). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 32.39%. Abbreviations: SpvO2: peripheral venous oxygen saturation. ScvO2: central venous oxygen saturation. c Peripheral and central venous oxygen saturation paired values, differences against means 3–4 h postoperatively. The Bland-Altman plot presents the differences against the means of peripheral and central venous oxygen saturation measurements obtained 3–4 h postoperatively. A total of 92 paired measurements were obtained. Bias was 16.49 (95% CI: 14.16–18.82, LoA: 21.20). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 26.83%. Abbreviations: SpvO2: peripheral venous oxygen saturation. ScvO2: central venous oxygen saturation

The trending ability of ScvO2 compared with SpvO2 in all measurements was investigated in a four-quadrant plot as demonstrated in Fig. 2. The association between central and peripheral venous oxygen saturation is presented in a scatter plot with a line of identity (Fig. 4a). The plot shows an 89% concordance with no zone of exclusion and 77% concordance including the zone of exclusion.

Fig. 2
figure 2

Trending ability of peripheral and central venous oxygen saturation at different time points. Four quadrant plot depicting trending ability of peripheral compared with central venous oxygen saturation between the different sampling points (preoperative, upon arrival to the ICU and 3–4 h postoperative). A total of 316 paired measurements were inserted in the four quadrant plot. The x axis represents the delta SpvO2 and the y-axis represents the delta ScvO2. Abbreviations: SpvO2: peripheral venous oxygen saturation. ScvO2: central venous oxygen saturation

Lactate

The mean value of central and peripheral lactate was 0.87 (±0.28) and 1.00 (±0.35) preoperatively, 1.71 (±1.15) and 1.89 (±1.22) on arrival to the ICU and 1.85 (±1.08) and 2.12 (±1.41) at 3–4 h postoperatively, respectively.

The Bland-Altman plots for lactate preoperatively are presented in Fig. 3ac. Bias was 0.14 mmol/L preoperatively (95% CI: 0.11–0.17, LoA: ±0.30), 0.16 mmol/L at arrival to ICU (95% CI: 0.09–0.23, LoA: ±0.70) and 0.23 mmol/L at 3–4 h postoperatively (95% CI: 0.11–0.35, LoA: ±0.50). The percentage error was 32.08% preoperatively, 38.88% at arrival to the ICU group and 25.18% at 3–4 h postoperatively. The association between central and peripheral lactate is presented in a scatter plot with a line of identity (Fig. 4b).

Fig. 3
figure 3

a Peripheral and central lactate paired values, differences against means preoperatively. The Bland-Altman plot presents the differences against the means of peripheral and central lactate measurements obtained preoperatively. A total of 111 paired measurements were obtained. Bias was 0.14 (95% CI: 0.11–0.17, LoA: 0.30). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 32.08%. b Peripheral and central lactate paired values, differences against means upon arrival to the ICU. The Bland-Altman plot presents the differences against the means of peripheral and central lactate measurements obtained upon arrival to the ICU. A total of 100 paired measurements were obtained. Bias was 0.16 (95% CI: 0.09–0.23, LoA: 0.70). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 38.88%. c Peripheral and central lactate paired values, differences against means 3–4 h postoperatively. The Bland-Altman plot presents the differences against the means of peripheral and central lactate measurements obtained 3–4 h postoperatively. A total of 87 paired measurements were obtained. Bias was 0.23 (95% CI: 0.11–0.35, LoA: 0.50). The upper and lower LoA are presented with +/− 1.96 SD (95% CI) (dashed lines). The percentage error was 25.18%

Fig. 4
figure 4

a The association between central and peripheral venous oxygen saturation. The scatter plot demonstrates the association between central and peripheral venous oxygen saturation measurements at all three time points. The solid line represents the line of identity and the dashed line represents the linear regression line. R2 = 0.12. b The association between central and peripheral lactate. The scatter plot demonstrates the association between central and peripheral lactate measurements at all three time points. The solid line represents the line of identity and the dashed line represents the linear regression line. R2 = 0.89

Discussion

We found a high bias between the measurements of SpvO2 and ScvO2 at all time-points. However, there was a moderate agreement with a percentage error of 22% to 32%. Furthermore, a moderate trending ability between measurements was found in the present study. A high saturation in peripheral measurements at baseline indicated arterialisation in some patients. Peripheral and central lactate levels showed excellent agreement at all time-points.

A reduction in venous oxygen saturation indicates an increased oxygen consumption/supply-ratio. In the absence of anaemia and arterial hypoxaemia, a low venous oxygen saturation reflects low cardiac output, which may be due to heart failure or obstruction of the circulation as in tamponade or hypovolaemia [3, 25]. Therefore, venous oxygen saturation may be a good indicator of impaired tissue oxygenation [1,2,3]. Low venous oxygen saturation has previously been associated with impaired outcome in surgical patients, and adequate measures of venous oxygen saturation may have a substantial role in management and treatment of critically ill patients [10].

Placement of a central or a pulmonary artery catheter carries some degree of risk and requires expertise to etablish [26]. Measurements of peripheral oxygen saturation and lactate are beneficial for patients without a central or a pulmonary artery catheter [21]. This would be valuable in an emergency or trauma setting avoiding unnecessary and time consuming application of a CVC in these patients in order to quickly identify patient hemodynamic. We found a high bias indicating a poor correlation between SpvO2 and ScvO2. However, differences in results between peripheral and central measurements were expected and the objective of the study was not to find a correlation but an association between measurements. Our results showed narrow LoA and a moderate trending ability between measurements, indicating that SpvO2 could be a less invasive measure to identify critically ill patients. One study has previously investigated the bias between SpvO2 and ScvO2 [21] however, the cut-off level for SpvO2 has not been estimated. Our results showed higher levels of saturation in SpvO2 compared to ScvO2. High levels of SpvO2 and ScvO2 at baseline, indicated arterialisation of the blood. As this was only seen in the measurements obtained at baseline and prior to CPB, this was most probably due to preoxygenation of patients prior to induction of anaesthesia. Arterial dissolved oxygen as a determinant of venous oxygen saturation has previously been described as a potential cause to misinterpretation of ScvO2 [27].

Peripheral lactate measurements have previously shown to have agreement with central venous [21] and arterial lactate measurements [28,29,30]. In accordance with a previous study, peripheral and central lactate showed excellent agreement and peripheral lactate has potential as a substitute for central lactate measures [21]. However, clinical studies are necessary to validate the use of peripheral lactate to identify patients at risk. The association between central and peripheral lactate is presented in a scatter plot with a line of identity (Fig. 3b). The linear regression line was R2 =0.89.

Limitations

Our results showed an acceptable LoA and a moderate trending ability, indicating an acceptable agreement between SpvO2 and ScvO2. The precision of the individual methods was not determined in this study and the exact value of an acceptable LOA of agreement could not be calculated. However, a percentage error of 22% to 32% is rated to be within the clinically acceptable [23]. Measurements of peripheral venous oxygen saturation and lactate could be valuable in different populations such as emergency or trauma settings, avoiding unnecessary and time consuming application of a central venous or a pulmonary artery catheter, and in order to quickly identify patient at risk and hemodynamics. Routinely all patients get a central and peripheral line prior to cardiac surgery at our department, which allowed us to investigate the feasibility of SpvO2 in this patient population. Additional studies are required to investigate whether these observations are reproducible in a different setting. The present study is limited by the fact that only a few patients demonstrated low cardiac output postoperatively and therefore, the clinical implication of SpvO2 in patient with low cardiac output is unknown. Surprisingly, a high SpvO2 was seen in many patients at baseline measurements. A simultaneously high ScvO2 indicated arterialisation in these patients. The risk of arterialisation at different ScvO2 values is unknown and should be further investigated. Other limitations to this study were differences in the peripheral intravenous site, which may have altered the results. Tourniquet application in some patients may have caused hemolysis in the peripheral blood samples. However, very few patients had permanence use of a tourniquet during peripheral blood sampling. Venous blood samples were obtained simultaneously from a central or a peripheral venous catheter however, delays may have occurred due to waste drawn.

Conclusions

To conclude, our result indicated that SpvO2 might be a valuable surrogate for ScvO2. As for peripheral lactate, measurements showed excellent agreement and is a reliable substitute for central lactate measurements. Additional studies are required to investigate the cut-off level for SpvO2 and to clarify whether observations are reproducible in patients with low cardiac output or in emergency settings.

Abbreviations

CI:

Confidence interval

CPB:

Cardiopulmonary bypass

CVC:

Central venous catheter

ICU:

Intensive Care Unit

LoA:

Limits of agreement

PE:

Percentage error

ScvO2 :

Central venous oxygen saturation

SmvO2 :

Mixed venous oxygen saturation

SpvO2 :

Peripheral venous oxygen saturation

References

  1. Walley KR. Use of central venous oxygen saturation to guide therapy. Am J Respir Crit Care Med. 2011;184:514–20.

    Article  PubMed  Google Scholar 

  2. Inomata S, Nishikawa T, Taguchi M. Continuous monitoring of mixed venous oxygen saturation for detecting alterations in cardiac output after discontinuation of cardiopulmonary bypass. Br J Anaesth. 1994;72:11–6.

    Article  PubMed  CAS  Google Scholar 

  3. Svenmarker S, Häggmark S, Östman M, Holmgren A, et al. Central venous oxygen saturation during cardiopulmonary bypass predicts 3-year survival. Interact Cardiovasc Thorac Surg. 2013;16:21–6.

    Article  PubMed  Google Scholar 

  4. Perz S, Uhlig T, Kohl M, Bredle DL, et al. Low and “supranormal” central venous oxygen saturation and markers of tissue hypoxia in cardiac surgery patients: a prospective observational study. Intensive Care Med. 2011;37:52–9.

    Article  PubMed  CAS  Google Scholar 

  5. Holm J, Håkanson E, Vánky F, Svedjeholm R. Mixed venous oxygen saturation predicts short- and long-term outcome after coronary artery bypass grafting surgery: a retrospective cohort analysis. Br J Anaesth. 2011;107:344–50.

    Article  PubMed  CAS  Google Scholar 

  6. Holm J, Håkanson RE, Vánky F, Svedjeholm R. Mixed venous oxygen saturation is a prognostic marker after surgery for aortic stenosis. Acta Anaesthesiol Scand. 2010;54:589–95.

    Article  PubMed  CAS  Google Scholar 

  7. Pölönen P, Ruokonen E, Hippeläinen M, Pöyhönen M, et al. A prospective, randomized study of goal-oriented hemodynamic therapy in cardiac surgical patients. Anesth Analg. 2000;90:1052–9.

    Article  PubMed  Google Scholar 

  8. Balzer F, Sander M, Simon M, Spies C, et al. High central venous saturation after cardiac surgery is associated with increased organ failure and long-term mortality: an observational cross-sectional study. Crit Care. 2015;19:168.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Svedjeholm R, Håkanson E, Szabó Z. Routine SvO2 measurement after CABG surgery with a surgically introduced pulmonary artery catheter. Eur J Cardio-thoracic Surg. 1999;16:450–7.

    Article  CAS  Google Scholar 

  10. Chemtob RA, Eskesen TG, Moeller-Soerensen H, Perner A, et al. Systematic review of the association of venous oxygenation and outcome in adult hospitalized patients. Acta Anaesthesiol Scand. 2016;60:1367–78.

    Article  PubMed  CAS  Google Scholar 

  11. Pope JV, Jones AE, Gaieski DF, Arnold RC, et al. Multicenter study of central venous oxygen saturation (ScvO2) as a predictor of mortality in patients with Sepsis. Ann Emerg Med. 2010;55:40–6.

    Article  PubMed  Google Scholar 

  12. Silva JM, Oliveira AMRR, de Morais SZ, de Araújo LS, et al. Influence of central venous oxygen saturation on in-hospital mortality of surgical patients. Rev Bras Anestesiol. 2010;60:593–602. 329–34

    Article  PubMed  Google Scholar 

  13. Bracht H, Hänggi M, Jeker B, Wegmüller N, et al. Incidence of low central venous oxygen saturation during unplanned admissions in a multidisciplinary intensive care unit: an observational study. Crit Care. 2007;11:R2.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Boulain T, Garot D, Vignon P, Lascarrou J-B, et al. Prevalence of low central venous oxygen saturation in the first hours of intensive care unit admission and associated mortality in septic shock patients: a prospective multicentre study. Crit Care. 2014;18:609.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Chung K-P, Chang H-T, Huang Y-T, Liao C-H, et al. Central venous oxygen saturation under non-Protocolized resuscitation is not related to survival in severe Sepsis or septic shock. Shock. 2012;38:584–91.

    Article  PubMed  CAS  Google Scholar 

  16. Textoris J, Fouché L, Wiramus S, Antonini F, et al. High central venous oxygen saturation in the latter stages of septic shock is associated with increased mortality. Crit Care. 2011;15:R176.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Rivers EP, Ander DS, Powell D. Central venous oxygen saturation monitoring in the critically ill patient. Curr Opin Crit Care. 2001;7:204–11.

    Article  PubMed  CAS  Google Scholar 

  18. Jones AE, Shapiro NI, Trzeciak S, Arnold RC, et al. Lactate clearance vs central venous. Jama. 2010;303:739–46.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  19. Heiselman D, Jeffrey Jones LC. Continuous Monitoring Of Mixed Venous Oxygen Saturation In Septic Shock. J Clin Monit. 1986;2:237–45.

    Article  PubMed  CAS  Google Scholar 

  20. Varpula M, Tallgren M, Saukkonen K, Voipio-Pulkki LM, et al. Hemodynamic variables related to outcome in septic shock. Intensive Care Med. 2005;31:1066–71.

    Article  PubMed  Google Scholar 

  21. Goyal N, Taylor AR, Rivers EP. Relationship between central and peripheral venous oxygen saturation and lactate levels: a prospective study. J Emerg Med. 2016;50:809–17.

    Article  PubMed  Google Scholar 

  22. Montenij LJ, Buhre WF, Jansen JR, Kruitwagen CL, et al. Methodology of method comparison studies evaluating the validity of cardiac output monitors: a stepwise approach and checklist. Br J Anaesth. 2016;116:750–8.

    Article  PubMed  CAS  Google Scholar 

  23. Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25:536–46.

    Article  PubMed  Google Scholar 

  24. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–10.

    Article  Google Scholar 

  25. Madsen P, Olesen HL, Klokker MSN. Peripheral venous oxygen saturation during head-up tilt induced hypovolaemic shock in humans. Scand J Clin Lab Invest. 1993;54:411–6.

    Article  Google Scholar 

  26. Carlbom DJ, Rubenfeld GD. Barriers to implementing protocol-based sepsis resuscitation in the emergency department—results of a national survey*. Crit Care Med. 2007;35:2525–32.

    Article  PubMed  Google Scholar 

  27. Legrand M, Vallée F, Mateo J, Payen D. Influence of arterial dissolved oxygen level on venous oxygen saturation: Don’t forget the PaO2! Shock. 2014;41:510–3.

    Article  PubMed  CAS  Google Scholar 

  28. Browning R, Datta D, Gray AJ, Graham C. Peripheral venous and arterial lactate agreement in septic patients in the emergency department: a pilot study. Eur J Emerg Med. 2014;21:139–41.

    Article  PubMed  Google Scholar 

  29. Nascente APM, Assunção M, Guedes CJ, Freitas FGR, Mazza BF, et al. Comparison of lactate values obtained from different sites and their clinical significance in patients with severe sepsis. Sao Paulo Med J. 2011;129:11–6.

    Article  PubMed  Google Scholar 

  30. Younger JG, Falk JL, Rothrock SG. Relationship between arterial and peripheral venous lactate levels. Acad Emerg Med. 1996;3:730–4.

    Article  PubMed  CAS  Google Scholar 

Download references

Funding

Departmental funding only.

Availability of data and materials

The datasets generated and/or analysed during the current study are available from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

RC collected, analysed and interpreted the data. RC wrote the original manuscript. HMS equally analysed and interpreted the data. HMS supervised, edited and revised the manuscript. Both authors read and approved the final manuscript.

Corresponding author

Correspondence to Raphaelle Avigael Chemtob.

Ethics declarations

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions