Table 2 Comparison between standard mechanical CPR and ACD CPR

From: Haemodynamic outcomes during piston-based mechanical CPR with or without active decompression in a porcine model of cardiac arrest

 

Standard CPR

mean ± SD

ACD-CPR

mean ± SD

Mean difference

(95% CI)

Standard CPR

median

(quartiles)

ACD-CPR

median

(quartiles)

p-value

Blood pressures (mmHg)

 Aortic pressure

 

55 (51, 70)

60 (51, 70)

0.86

 Right atrial pressure

54 ± 16

55 ± 16

1.0 (− 4.0, 6.0)

 

0.66

 Coronary perfusion pressure

2.2 ± 22

0.6 ± 20

− 2.7 (− 10, 4.7)

 

0.42

 Oesophageal pressure

57 ± 38

60 ± 39

2.6 (− 2.6, 7.8)

 

0.28

 Intracranial pressure

24 ± 6.4

23 ± 6.7

−0.7 (− 1.6, 0.1)

 

0.09

Blood flow

 Cerebral flow (AU)

253 ± 268

297 ± 264

44 (14, 74)

 

0.01

 Carotid artery flow (ml/min)

 

83 (57, 94)

97 (70, 106)

< 0.01

Manual measurements

 Cardiac output (l/min)

 

1.1 (0.8, 1.5)

1.5 (1.1, 1.7)

< 0.01

 End tidal CO2 (kPa)

2.2 ± 0.8

2.5 ± 1.0

0.3 (− 0.1, 0.7)

  

0.15

Blood gases (kPa)

 pH - Arterial

 

7.3 (7.2, 7.4)

7.3 (7.2, 7.4)

0.33

 pCO2 - Arterial

 

5.6 (4.9, 8.0)

5.4 (4.4, 8.1)

0.89

 pO2 - Arterial

22 ± 20

25 ± 21

2.5 (−6.2, 11)

  

0.52

 pH - Venous

 

7.1 (7.1, 7.2)

7.2 (7,1, 7.2)

0.09

 pCO2 - Venous

11 ± 1.9

11 ± 1.9

−0.3 (− 0.5, 0.0)

 

0.06

 pO2 - Venous

 

3.1 (3.0, 3.6)

3.3 (3.1, 3.6)

0.11