Table 3 Radiation dose during the different series investigated

From: Maintaining immobilisation devices on trauma patients during CT: a feasibility study

  

CTDIvol [mGy]

DLP [mGycm]

Case

Device

Heada

Neck

Chest

Abdomen

Heada

Neck

Chest

Abdomen

1

No device

50.2

11.2

2.7

5.7

929

253

96

240

2

Vacuum mattress 1

52.3

11.3

3.1

7.1

963

252

108

301

3

Vacuum mattress 2

54.3

12.4

3.7

6.7

983

299

135

297

4

Vacuum mattress 3

59.3

11.9

3.5

7.2

1094

260

123

293

5

Spine board 1

61.2

11.4

3.1

7.0

1111

252

113

291

6

Spine board 2

50.0

11.3

3.1

6.6

918

250

111

275

7

Trauma transfer

53.8

11.1

3.1

6.5

991

252

108

273

8

Spine board 1 with IMAR

61.2

11.4

3.1

7.0

1111

252

113

291

  1. a 16 cm CTDI phantom, all others 32 cm phantom