Table 4 Comparison of different methods of measurement ("tymp" as reference value)

From: Temperature measurements in trauma patients: is the ear the key to the core?

Methods n Bias SD LOA p-value
Bladder 111 0.42 0.48 - 0.52, 1.35 0.00
Temporal 20 0.43 1.31 - 2.15, 3.0 0.16
IRtymp 23 0.45 0.62 - 0.77, 1.66 0.00
Oesophageal 5 0.78 0.31 0.17, 1.39 0.00
Naso 40 0.03 0.35 - 0.65, 0.71 0.57
Bladder vs IRtymp 15 -0.13 0.36 - 0.83, 0.57 0.17