Skip to main content

Table 1 Key scanning parameters in the experimental and the control group

From: Dose reduction in whole-body computed tomography of multiple injuries (DoReMI): protocol for a prospective cohort study

 

Standard dose

Low dose

 

128-row MDCT pan-scan

128-row MDCT pan-scan + iDose

Detector z coverage, mm

40

40

kVp

120

120

Exposure, mAs

180

90

iDose Level

off

4 (6)

Recon filter

B

B (C)

CTDI, mGy

11.9

5.9