From: End-tidal carbon dioxide monitoring during bag valve ventilation: the use of a new portable device
Minutes Bag valve Ventilation | Time point in article | Bias (Mean difference) | 95% confidence interval for bias | Limits of agreement (Lower LoA, Upper LoA) | 95% confidence interval for LoA (CI Lower LoA; CI Upper LoA) |
---|---|---|---|---|---|
5 | 1 | -1.48 | -1.80, -1.15 | -2.92, -0.03 | -3.56, -2.28; -0.68, 0.61 |
10 | 2 | -1.37 | -1.66, -1.08 | -2.64, -0.09 | 3.24, -2.04; -0.70, 0.51 |
15 | 3 | -1.62 | -1.93, -1.31 | -2.99, -0.24 | -3.62, -2.37; -0.87, 0.38 |
20 | 4 | -1.46 | -1.85, -1.08 | -3.17, 0.25 | -3.88, -2.47; -0.46, 0.95 |